Biomedical Devices
How the body's electricity becomes a clean ECG: electrodes, patient safety, amplifiers, filters, power and sampling, ending with a real ECG board.
THE SYLLABUS
I
0 / 5
▶
Why the Body Makes Electricity
Every cell keeps a voltage across its membrane by holding ions apart: about −70 mV in a nerve and about −90 mV in heart muscle. When sodium channels open, Na+ rushes in and the inside swings positive; when potassium flows out, it returns. That action potential swings about 100 mV. An electrode on the skin doesn't see one cell: it sees a faint, smeared sum of millions, about 1 mV for the heart.
Start
🔒
ECG, EMG and EEG
The body's three main signals differ by size and speed. The ECG from the heart is about 0.5 to 4 mV, with content from 0.05 to 150 Hz. Surface EMG from muscles is 0.1 to 5 mV but fast, 20 to 500 Hz. The EEG from the brain is only 10 to 100 µV, from 0.5 to 100 Hz: ten to a hundred times smaller than the ECG. Those numbers decide how each one has to be amplified and filtered.
Premium
🔒
Signal vs. Noise
A 1 mV heartbeat has to be found in a room full of interference: 60 Hz hum from mains cords and lamps, bursts from a phone, the baseline jumping when the patient moves, and fast muscle tremor at 20 to 500 Hz. Several are bigger than the ECG itself. Signal-to-noise ratio says how badly: SNR in dB = 20·log10(V_signal / V_noise), so 1 mV under 10 mV of hum is −20 dB. Mains pickup on the body can even reach volts.
Premium
🔒
From Signal to Requirements
A front end's spec comes straight from the signal's numbers. Gain is the output span over the input span: a 1 mV ECG brought up to 0.5 V needs ×500, about 54 dB, which is what the course's ECG board does. The band edges come from the signal's bandwidth: a high-pass to remove drift, a low-pass to stop fast noise. And because mains hum can put about a volt on both inputs, the amplifier needs a high common-mode rejection. Gain is only one of the requirements.
Premium
🔒
Level Review
Level 1 review: signal sizes and speeds, resting potential, gain and dB, SNR, volts per division, two noise traces, and two items from Seeing Circuits.